10.3. Care, transportation, and storage
‚""""""
The device may be cleaned with alcohol. Attention: When handling combustible fl uids do not
smoke or operate with open fl ame (e.g. candles, etc.)
‚""""""
Stains and massive soiling may be removed with a mild detergent.
‚""""""
Do not emerge the device into liquids or subject it to large amounts of water.
‚""""""
For save transportation always insert the device into the carrying case.
‚""""""
If the device is not used for a prolonged period of time, remove the batteries from the battery
compartment (acid could leak and damage the device). Store the device and all accessories
cool and dry in the carrying case.
10.4. Safety-technical control measures, STC
According to the medical products operator’s regulation (MPBetreibV), § 6 (safety-technical
control measures), the user is obliged to have regular safety-technical controls conducted.
hofmann gmbh prescribes these controls in accordance to MPBetreibV § 6 in a 24-months turn.
The safety-technical control may only be delegated to persons, who are able to conduct controls in
an orderly fashion based on their education, knowledge, and experience achieved by practical work,
and who also have suitable measuring and proofi ng tools available.
Standards for safety-technical control are to be requested from the quality management agent of
hofmann gmbh.
Prior to each application of the sanowell Life Tens device the following controls are to be
conducted:
1. Inspect device for surface damage:
‚ """""
Deformation of casing?
‚ """""
Damaged or defect electrode sockets?
2. Inspect device for damaged control elements:
‚ """""
Imprints still legible?