background image

OPERATOR’S MANUAL 

VER1.0C615 

Fingertip Pulse Oximeter

 

 

General Description   

Oxygen Saturation is a percentage of Oxyhemoglobin (HbO2) capacity, compounded 
with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it 
is consistency of Oxyhemoglobin in blood. It is a very important parameter for the 
Respiratory Circulation System. Many respiratory diseases can result in oxygen 
saturation being lowered in human blood. Additionally, the following factors can 
reduce oxygen saturation: Automatic regulation of organ dysfunction caused by 
Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical 
examinations. That situation might result in light-headedness, asthenia, and vomiting. 
Therefore, it is very important to know the oxygen saturation of a patient so that 
doctors can find problems in a timely manner. 

The fingertip pulse Oximeter features small size, low power consumption, convenient 
operation and portability. It is only necessary for a patient to put one of his fingers into 
the fingertip photoelectric sensor for diagnosis, and a display screen will show oxygen 
saturation. It has been proven in clinical experiments that it also features high 
precision and repeatability.

 

Measurement Principle 

Principle of the Oximeter is as follows: A mathematical formula is established making 
use of Lambert Beer Law according to Spectrum Absorption Characteristics of 
Reductive hemoglobin(RHb) and Oxyhemoglobin (HbO2) in glow and near-infrared 
zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection 
Technology is adopted in accordance with Capacity Pulse Scanning and Recording 
Technology, so that two beams of different wavelength of lights (660nm glow and 
940nm near infrared light) can be focused onto a human nail tip through a clamping 
finger-type sensor. A measured signal obtained by a photosensitive element, will be 
shown on the Oximeter’s display through process in electronic circuits and 
microprocessor shown on the Oximeter’s display through electronic circuits and a 
microprocessor. 

Diagram of Operation Principle 

1.  Red and Infrared-ray Emission Tube 
2.  Red and Infrared-ray Receipt Tube 

Product Operation Scope 

Fingertip Pulse Oximeter is a portable non-invasive, spot-check, oxygen saturation of 
arterial hemoglobin (SpO

2

) and pulse rate of adult and pediatric patients at home, and 

hospital (including clinical use in internal medicine/surgery, anesthesia, intensive care 
and etc). The device is not intended for continuously monitoring. 

The PULE OXIMETER requires no routine calibration or maintenance other than 
replacement of batteries.

 

Technical Specifications 

1. Display: 

LED 

2. SpO

2

Measurement range

70

99% 

Accuracy

 70%

99%,  ±3%;  <70% no definition    

3. Pulse Rate: 

Measure range

30~235 bpm 

Accuracy

30~100bpm ±2bpm; 101~235bpm, ±2% 

Pulse Intensity: Bar graph Indicator   

4. Power Requirements:

   

Two AAA alkaline Batteries 
Power consumption: Less than 30mA   

Low power indication:

 

 

Battery Life: Two AAA 1.5V, 800mAh alkaline batteries could be continuously 
operated as long as 30 hours.   

5. Dimension: 

Length: 58mm 
Width: 32mm 
Height: 37mm 
Weight: 33g (without batteries) 

6. Environment Requirements: 

Operation Temperature: 5

40

 

Storage Temperature: -20

55

 

Ambient Humidity: 

80%, no condensation in operation;

 

93%, no condensation in storage

 

7.Measurement Performance in Low Perfusion Condition: 

The pulse wave is 

available without failure when the simulation pulse wave amplitude is at 0.6% using 
the test equipment (BIO-TEK INDEX Pulse Oximeter Tester).   

8.Interference Resistance Capacity against Ambient Light:

 Device works 

normally when mixed noise produced by BIO-TEK INDEX Pulse Oximeter Tester.

 

                                                                                                                             

                                                                                                                                                                   

Precautions for use 

Do not use the pulse oximeter in an MRI or CT environment 

Do not use the pulse oximeter in situations where alarms are required. The 
device has no alarms. 

Explosion hazard:

 Do not use the pulse oximeter in an explosive atmosphere. 

The pulse oximeter is intended only as an adjunct in patient assessment. It 
must be used in conjunction with other methods of assessing clinical signs and 
symptoms. 

Check the pulse oximeter sensor application site frequently to determine the 
positioning of the sensor and circulation and skin sensitivity of the patient.   

Do not stretch the adhesive tape while applying the pulse oximeter sensor. This 
may cause inaccurate readings or skin blisters. 

Before use, carefully read the manual. 

The pulse oximeter has no SpO

2

 alarms; it is not for continuous monitoring, as 

indicated by the symbol.   

Prolonged use or the patient’s condition may require changing the sensor site 
periodically. Change sensor site and check skin integrity, circulatory status, and 
correct alignment at least every 4 hours. 

10 

Inaccurate measurements may be caused by autoclaving, ethylene oxide 
sterilizing, or immersing the sensors in liquid may cause inaccurate readings. 

11 

Significant levels of dysfunctional hemoglobins(such as carbonxy-hemoglobin 
or methemoglobin). 

12 

Intravascular dyes such as indocyanine green or methylene blue 

13 SpO

2

 measurements may be adversely affected in the presence of high 

ambient light. Shield the sensor area (with a surgical towel, or direct sunlight, 
for example) if necessary. 

14 

Excessive patient movement may cause accurate readings. 

15 

Venous pulsations may cause accurate readings. 

16 

Placement of a sensor on an extremity with a blood pressure cuff, arterial 
catheter, or intravascular line may cause accurate readings. 

17 

The patient has hypotension, severe vasoconstriction, severe anemia, or 
hypothermia. 

18 

The patient is in cardiac arrest or in shock. 

19 

Fingernail polish or false fingernails may cause inaccurate SpO

2

 readings. 

Follow local ordinances and recycling instructions regarding disposal or 
recycling of the device and device components, including batteries. 

Product Properties 

1.  Easy to use.  

2.    Small in volume, light in weight. 

3.  Lower power consumption.  

4.    Low voltage warning will be indicated in visual window when battery voltage is low. 

5.    The product will automatically be powered off when no signal is in the product for 

longer than 8 seconds.   

Operation Instructions 

1.    Install two AAA batteries into battery cassette, and then close its cover. 

2.    Open the clamp as shown in the picture below. 

3.    Insert one finger into the flexible fingertip of the Oximeter fully. 

4.    Press the power switch once on the front panel. 

5.    The patient’s finger should remain still while the Oximeter is working.   

6.    Read corresponding data from display screen. 

When you insert a finger into the Oximeter, your nail surface must be upward. 

 

NOTE:

 Please use the medical alcohol to clean the rubber touching the finger inside of 

Oximeter, and clean the test sensor using alcohol before and after operation. (The 
rubber inside of the Oximeter is medical rubber, which has no toxin and no harm to the 
skin of human being). 

Product Accessories 

1. One strap 

2. Two batteries 

3. One user’s manual

 

Strap Installation 

1.    Thread thinner end of the strap through the loop. 

2.    Thread thicker end of the strap through the threaded end before pulling it tightly. 

 

Brief Description of Front Panel 

Reviews: