— 13 —
Remarks:
s
For capacitance measurement,
press the button to
cancel the indicated value
(00.00 nF) before connecting a
capacitor.
s
Only the auto range is available
for the capacitance measuring
function.
s
The indication may not become
stable due to influence of
surrounding noises or stray
capacitance of the test leads.
Function
Max. Rated Input
Range
CAP ( )
100.0 μF
50.00 nF, 500.0 nF, 5.000 μF,
50.00 μF, 100.0 μF (Auto range)
5-5 Capacitance Measurement ( )
Cancel
indicated
value prior to
measurement.
Remarks:
s
Input sensitivity: 3 Vrms or over.
s
Zero cross (+ potential
- potential
+ potential) frequencies can be
measured. Frequencies of + potential only or – potential only such as
logic pulses cannot be measured.
s
Frequencies less than 1 Hz cannot be measured.
s
When the Hz function is used, the data hold and relative functions
cannot be used.
Never apply a voltage to the measuring terminals.
WARNING
1. Remove electric charge in the capacitor prior to measurement.
2. Because this meter applies a current to the capacitor to measure, it is
not suitable for measurement of electrolytic capacitors having a large
leak current as a large error will occur.
3. For capacitors having large capacitance, measurement takesa longer time.
CAUTION