− 43 −
⑦
Read the value on the display as follows.
DC20A
: multiplying by 0.1
AC20A
: multiplying by 100
DC200A : multiplying by 1
AC200A : multiplying by 1000
⑧
After measurement, open the clamp part and release clamp probe
from the electric wire.
⑨
Turn the function switch of multimeter and selector knob of
clamp probe to the position of OFF.
5-8-3 DC Clamp Probe (CL33DC) Maximum measurement value DC300A
<Measurement procedure> (See Fig 17, page 20)
①
Connect the black plug of the current probe to the COM
input
terminal and the red plug to the
input terminal.
②
Set the function switch at "V
/
" function.
③
Select the select switch at "
" (DC) and set the 400mV range
with the range hold switch.
④
Select either 30A or 300A with selector knob of clamp probe.
⑤
Open the clamp part, have electric line (one line) clamped.
⑥
Read the value on the display in A unit and when current probe
of the 30A range after multiplying indicated value by 0.1, and the
300A range after multiplying indicated value by 1.
⑦
After measurement, open the clamp part and release clamp probe
from the electric wire.
⑧
Turn the function switch of multimeter and selector knob of
clamp probe to the position of OFF.
5-8-4 Temperature probe (T-300PC) Measurement range 300˚C from -50˚C
<Measurement procedure> (See Fig 18, page 21)
①
Connect the black plug to COM
input terminal and the red
plug to
input terminal.
②
Set the function switch at "
Ω
/ / "
function.
③
Select the select switch at "
Ω
" and set the 4k
Ω
range with the
range hold switch.
④
Apply the sensor to an object to measure.
⑤
Read the value on the display.
(When measuring temperature, please neglect the value of k
Ω
indicated on LCD of multimeter. The exact value of measured
teperature is shown only on the window of PCLink.)
⑥
After measurement, release the sensor from the object measured.
⑦
Turn the function switch to the position of OFF.
Summary of Contents for PC20
Page 1: ...PC20 DIGITAL MULTIMETER 取扱説明書 INSTRUCTION MANUAL ...
Page 2: ......