10
REFRIGERATOR COMPONENTS
Control panel and keypad
1. Buzzer stop key (BUZZER):
To silence the audible alarm, press this key.
2. Temperature display key (UPPER/LOWER):
By pressing this key, the displayed temperature is
changed; upper monitor bottle temperature, lower monitor bottle temperature, and average temperature
of upper and lower monitor bottles are displayed in sequence.
3. Alarm test key (ALARM TEST):
To check the operation of alarm buzzer, lamp, and remote alarm
terminal.
4. Calibration key (CAL):
To calibrate the temperature. Normally this key is not used.
5. Digit shift key ( ):
Pressing this key in the CAL mode will cause the changeable digits to move.
6. Numerical value shift key ( ):
Pressing this key in the CAL mode will cause the numerical value
to change.
7. Door alarm lamp (DOOR):
This lamp is lit when the door is ajar or open.
8. Alarm lamp (ALARM):
This lamp flashes during an alarm condition.
9, 10. Temperature display monitor lamp (UPPER, LOWER):
Lamp 9 is lit when upper monitor bottle
temperature is displayed. Lamp 10 is lit when lower monitor bottle temperature is displayed.
Both 9 and 10 are lit when the average temperature is displayed.
11. Digital temperature indicator:
This indicator shows the upper or lower monitor bottle temperature
or average temperature.
1
2
3
4
11
8
7
5
6
9
10