![background image](http://html.mh-extra.com/html/sanyo/mco-19m/mco-19m_service-manual_361500026.webp)
- 23 -
Operational specifications of LCD panel
㻌
㻌
1. Display
descriptions
<Top Screen>
Turn the power switch on and Top Screen is displayed.
䣖 䣧 䣯 䣲
䢵 䢹 䢰 䢲 䤑䣅
䣅 䣑 䢴
䢷 䢰 䢲 䢧
䣃
䣄
䣑 䢴
䢷 䢰 䢲 䢧
䣃
䣄
䣕 䣶 䣣 䣶 䣷 䣵
䣑 䣍
䢷 䢰 䢲
䢷 䢰 䢲
䢵 䢹 䢰 䢲
z
Setting temperature: Digits in the right side of “Temp”
The digits are displayed in reverse video when heater is active.
z
Setting CO2 density: Digits in the right side of “CO2”
The digits are displayed in reverse video when CO2 valve is active.
z
Setting O2 density: Digits in the right side of “O2”
㻌
The digits are displayed in reverse video when O2/N2 valve is active.
z
Current temperature: Large digits in the left side
z
Current CO2 density: Large digits in the center
z
Current O2 density: Large digits in the right side
z
Status field: Status of UV lamp, humidifying water and door
z
Message field: Message is displayed when an apparatus is in low humidifying water level, door open,
high/low temperature alarm, high/low CO2 density alarm and high/low O2 density alarm.
When an apparatus is in normal condition, “OK” is displayed.
z
CO2 Auto changer display: “A” or “B” in the right side of setting CO2 density is displayed when an optional
Automatic CO2 cylinder changeover system (MCO-21GC) is installed.
Either “A” or “B” displayed in reverse video means that alternative CO2 cylinder
between A and B is current using.
When a current used cylinder becomes empty, digits in reverse video starts blinking.
In Top Screen, press right shift key (
䊻
) for 3 seconds to switch valve control from A to B.
In Top Screen, press left shift key (
䊹
) for 3 seconds to swtich valve control from B to A.
z
N2/O2 Auto changer display: Either “A” or “B” displayed in reverse video in the right side of setting O2 density
means the alternative O2 cylinder is current using.
When a current using cylinder becomes empty, digits in reverse video starts blinkking.
In Top Screen, press down key (
䊼
) for 3 seconds to switch valve control from A to B.
In Top Scrreen, press up key (
䊺
) for 3 seconds to switch valve control from B to A.
Summary of Contents for MCO-19M
Page 7: ... 4 Dimensions Power cord ...
Page 16: ...Wiring diagram 13 ...
Page 17: ...Circuit diagram Main PCB 14 ...
Page 18: ... LCD PCB 15 ...
Page 110: ...MCO 19M UVH MCO 19M UV MCO 19M Multi Gas Incubator INSTRUCTION MANUAL 107 ...
Page 112: ...CONTENTS SPECIFICATIONS P 74 PERFORMANCE P 75 SAFETY CHECK SHEET P 76 109 2 ...
Page 183: ...Fig A Stacking plate B Stacking plate A Protective sticker Front panel Hook Front 180 73 ...
Page 187: ...SANYO Electric Co Ltd Printed in Japan DC3186 150B ...