background image

Specifications

*Appearance and specifications are subject to change without notice.
*Cooling performance is indicated by the temperature reached at the center of the freezer (at ambient temperature of 30

°

C with no load). 

In order to use the freezer at a stable temperature for a long time, it is recommended that the temperature be set to at least 5

°

C higher 

than the indicated lowest temperature. In addition, depending on the usage conditions, it may not be possible to reach the indicated 
lowest temperature.

HFC

Model 
LCO

2

 backup system

LN

2

 backup system

Backup system mounting kit
Circular recorder

Chart paper
Ink pen

Slip chart recorder

Chart paper
Ink pen

Recorder mounting bracket

Sensor cover
Storage racks
3 drawers (for lower chamber)
SANYO DAQ (Data Acquisition)
system

MDF-C8V

CVK-UB4

MDF-UBK

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MTR-C8


MTR-480

MTR-2000

MDF-U32V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

 

(for MTR-85H)

MDF-30SC

MDF-30R

MTR-480

MTR-2000

MDF-U53V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MDF-70SC

MDF-50R

MTR-480

MTR-2000

MDF-U73V

CVK-UB2

CVK-UBN2

MTR-G85

RP-G85

PG-R

MTR-85H

RP-85

DF-38FP

MDF-S3085

(for MTR-85H)

MDF-70SC

MTR-480

MTR-2000

MDF-C2156VAN

Equipped




MTR-155H

RP-155

DF-38FP

MDF-S30150

 

(for MTR-155H)



MTR-480

MTR-2000

Options

Distributed by:

We design and manufacture products in accordance with ISO9001 & 14001 requirements. 
RoHS (European Restriction of Hazardous Substances directives) compliant

• All data shown here are measured based on our company standards. If detailed 

data is required, please contact our company.

• If product inspection records are required, please contact our company. (There 

will be a charge for the records.) Please note that the records do not include tem-
perature adjustment range or data on temperature distribution inside the unit.

Performance Data

40

20

0

–20

–120

–40
–60
–80

–100

0

7

6

5

2

8

3

4

1

T

e

mperature (˚C)

T

e

mperature (˚C)

Time (hours)

5

25

20

15

10

30

35

20

10

15

5

T

ime

 (

hour

s

)

T

im

e

 (hours)

Ambient temperature (˚C)

Ambient temperature (˚C)

0

20

10

30

40

20

25

10

15

5

Time (hours)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

2

20

4

6

8

10

18

16

12 14

Time (hours)

T

ime (hours)

Ambient temperature (˚C)

T

e

mperature (˚C)

5

25

20

15

10

30

35

20

10

15

5

Ambient temp. and maintenance 
of –70

°

C by injection of LCO

2

 (30kg)

5

25

20

15

10

30

35

20

10

15

5

T

ime (hours)

Ambient temperature (˚C)

5

25

20

15

10

30

35

10

15

20

5

T

ime (hours)

Ambient temperature (˚C)

MDF-C2156VAN

LN

2

 backup system 

operation characteristics

MDF-C8V

MDF-U32V

Central air temp. in cabinet
Ambient temp. 30

°

C, No-load

Ambient temp. and maintenance 
of –70

°

C by injection of LCO

2

 (30kg)

Central air temp. in cabinet
Ambient temp. 30

°

C, No-load

Ambient temp. and maintenance 
of –70

°

C by injection of LCO

2

 (30kg)

Central air temp. in cabinet
Ambient temp. 30

°

C, No-load

Ambient temp. and maintenance 
of –135

°

C by injection of LN

2

 (50kg)

Central air temp. in cabinet
Ambient temp. 30

°

C, No-load

Ambient temp. and maintenance 
of –70

°

C by injection of LCO

2

 (30kg)

Central air temp. in cabinet
Ambient temp. 30

°

C, No-load

MDF-U73V

MDF-U53V

T

e

mperature (˚C)

Time (hours)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

Time (hours)

T

e

mperature (˚C)

40
20

0

–20

–120

–40

–140

–60

–160

–80

–100

0

12

6

5

4

3

2

10

1

11

9

8

7

CO

2

 backup system 

operation characteristics

Pull-up characteristics 
during power failure

Pull-down characteristics

Further details regarding our products can be accessed at

 http://www.sanyo.co.jp/cmg/biomedical

Model

Temperature range

Maximum cooling performance

Exterior dimensions (W x D x H)

Interior dimensions (W x D x H)

Net weight

Effective capacity

Shelves

Access port

Compressor

Refrigerant

Alarm system

Remote alarm contact

Accessories

Model

Temperature range

Maximum cooling performance

Exterior dimensions (W x D x H)

Interior dimensions (W x D x H)

Net weight

Effective capacity

Access port

Compressor

Refrigerant

Alarm system

Remote alarm contact

Accessories

MDF-C8V

-60 to -80

°

C (1

°

C increments)

-80

°

C (Ambient temp. 30

°

C)

550 x 685 x 945 mm (21.6” x 27.0” x 37.2”)

405 x 490 x 425 mm (15.9” x 19.3” x 16.7”)

Approx. 67 kg (148lbs.)

84 Liters (3.0 cu.ft.)

17 mm diameter, 2 locations backside, bottom

Rotary type, Output: 400W

High/Low temp. alarm (

±

5

°

C to 

±

20

°

C),

Power failure alarm, Remote alarm contact: 

DC30V, 2A

1 set of keys, 1 scraper

MDF-U73V

1010 x 875 x 2010mm

(39.8” x 34.4” x 79.1”)

870 x 600 x 1400mm

(34.2” x 23.6” x 55.1”)

Approx. 350kg (772 lbs.)

728 Liters (25.7 cu.ft.)

MDF-U53V

–50˚C to –86˚C (1˚C increments)

–86˚C (Ambient temp. 30˚C)

770 x 875 x 1990mm

(30.3” x 34.4” x 78.3”)

630 x 600 x 1380mm

(24.8” x 23.6” x 54.3”)

Approx. 299kg (659 lbs.)

519 Liters (18.3 cu.ft.)

Stainless steel, 3 shelves

17mm diameter, 3 locations (back, bottom left/right corners)

Hermetic type, Output: 

1100 W (high stage side), 1100 W (low stage side)

High/low temperature, Power failure, Filter check,

Self diagnostics, Door check,

Part replacement nortification

Allowable contact capacity: DC 30V, 2A

1 set of keys, 1 scraper

MDF-U32V

670 x 867 x 1860mm 

(26.4” x 34.1” x 73.2”)

490 x 600 x 1140mm

(19.3” x 23.6” x 44.9”)

Approx. 258kg (569 lbs.)

333 Liters (11.8 cu.ft.)

Hermetic type, Output: 

400 W (high stage side),

600 W (low stage side)

Temperature, Power failure, 

Filter check, Self diagnostics, 

Remote alarm contact

MDF-C2156VAN

-125

°

C to -150

°

C (1

°

C increments)

-150

°

C (Ambient Temp. 30

°

C)

1730 x 765 x 1010 mm (68.1” x 30.1” x 39.8”)

760 x 495 x 615 mm (29.9” x 19.5” x 24.2”)

Approx. 319 kg (703lbs.)

231 Liters (8.2 cu.ft.)

40 mm diameter, 1 location

Hermetic type, Output: 1100W (high stage 

side), 1100W (low stage side)

HFC

High/Low temperature, Power failure,

Filter check, Self diagnosis, Door check,

Part replacement notification

Allowable contact capacity, DC 30V, 2A

1 set of keys, 1 scraper, Connecting tube for 

LN

2

 backup system

SANYO Electric Co., Ltd.

Biomedical Division

©2008 SANYO Printed in Japan 2008.3 MA

SHR119

Summary of Contents for MDF-C2156VAN

Page 1: ...Ultra Low Temperature V I P Series Freezers Integrated Ultra Low Temperature Storage Solutions for Biosafety Security and Performance ...

Page 2: ...essor controls and high performance refrigeration achieve the most efficient storage capacity in the industry Exterior Dimensions Unit mm inch Height Depth Width The ultra low temperature freezer VIP series lineup meets diversifying needs for ultra low temperature storage including storage of large quantities of biological genetic material quality control environmental consciousness etc ...

Page 3: ...rg discovered DNA polymerase Khorana succeeded in synthesis of active DNA Cohen Boyer established gene recombination technology Dangerousness of gene engineering study was discussed in Asilomar conference Successful human insulin production in E coli Mullis developed PCR method Human Genome Project began Human Genome Project completed History of SANYO Ultra Low Temperature Freezers 40 C Freezer 80...

Page 4: ...proof To prevent unauthorised access to freezer settings a key lock func tion is provided that blocks changes being made through the keypad Delay start On power failure recovery typically every unit in the lab will try to start up simultaneously This can create a significant drain on the electricity supply and a drop in voltage that may cause the breaker to trip again SANYO s exclusive compressor ...

Page 5: ...ting and operation Provides large capacity in a limited space and helps to meet constantly increasing sample storage needs Space saving design enables effective use of valuable laboratory space Innovative control panel A graphic LCD display panel with pop up menu function allows simple intuitive operation and provides detailed information of freezer status Chest freezer MDF C2156VAN for cryogenic ...

Page 6: ...temp 30 C No load Ambient temp and maintenance of 135 C by injection of LN2 50kg Central air temp in cabinet Ambient temp 30 C No load Ambient temp and maintenance of 70 C by injection of LCO2 30kg Central air temp in cabinet Ambient temp 30 C No load MDF U73V MDF U53V Temperature C Time hours 40 20 0 20 120 40 140 60 160 80 100 0 12 6 5 4 3 2 10 1 11 9 8 7 40 20 0 20 120 40 140 60 160 80 100 0 12...

Reviews: