background image

   

10      

Twin Guard Series

 -86°C Ultra-Low Freezers    

www.twinguardseries.com

Twin Guard Series

 Features

Product Features

1.  SANYO patented V.I.P. PLUS™ vacuum 

insulation panel cabinet construction 
for high insulating value and increased 
interior volume in the same footprint. 
The V.I.P. PLUS™ minimizes energy 
transfer to and from the ultra-low 
temperature interior. The composite 
construction, complete with reflective 
barrier film and structural closed-cell 
foam, is used on all walls and the 
outer door.  
 
This advanced insulation technology 
offers structural stability to eliminate 
distortion, and inhibits moisture accu-
mulation that can lead to icing. Aggre-
gate insulation efficiency minimizes 
compressor cycle run-time to lower 
energy costs.

2.  Easy-In/Easy-Out™ door latch for 

smooth, one-handed operation,  
positive seal against gasket. Key  
lock provision standard.

3.  Universal keyed door lock offers  

added security.

4.  Integrated, microprocessor-based con-

trol system and LCD display informa-
tion center includes comprehensive 
setpoint, alarm, monitoring, diagnostic 
and communications functions.

5.  Circular-chart temperature recorder 

(optional) mounts easily in pre-engi-
neered mounting space.

6.  Insulated and gasketed inner doors 

seal inside to offer additional protec-
tion, improve uniformity. Inner door 
latches are standard.

7.  New, air-cooled Super Condenser 

energy transfer technology maintains 
optimum condenser air flow and 
eliminates the need for an air filter or 
periodic cleaning and maintenance 
required by other manufacturers.

8.  High impact, recessed casters and 

leveling feet simplify installation.

9.  Seismic restraints secure cabinet  

to building superstructure to meet 
local codes.

10.  New generation SANYO designed 

Cool Safe™ compressors are specifi-
cally designed for one-compressor, 
autocascade applications.

11.  Twin Guard Series™ redundant refrig-

eration -86 Dual°Cool™ circuits offer 
two systems in one, each providing 
back-up for the other, or working 
together in EcoMode™ operation for 
additional energy savings.

12.  Multiple access ports permit inser-

tion of independent probes, instru-
mentation or liquid nitrogen or liquid 
CO

2

 back-up injectors.

13.  CFC-refrigerants are highly efficient, envi-

ronmentally safe, non-ozone depleting.

14.  Internal voltage and power management 

systems assure component protection 
over wide voltage ranges.

15.  A vacuum relief valve is mounted in the 

left wall.

16.  Remote alarm contacts and optional 

communication port available; see Acces-
sories.

17.  Vacuum relief valve

Patented SANYO refrigerants are non-ozone depleting, 
non-flammable and environmentally safe in compliance 
with the Montreal Protocol.

SANYO V.I.P. PLUS

 Series freezers offer high-density 

storage that effectively reduces the volumetric unit 
costs of ultra-low storage.

Summary of Contents for MDF-U500VXC

Page 1: ...es www sanyobiomedical com GREEN product Features SANYO 86 Dual Cool ultra low freezer is designed for 86 C storage of high value biologicals Includes integrated LCD performance monitor and digital controller for comprehensive system management 86 C Ultra Low Freezers MDF U500VXC MDF U700VXC ...

Page 2: ...e Energy SAVE Air Filter VIP Plus Vacuum Insulation Panel VIP Filterless Design Insulated Inner Doors ICD Digital Display Remote Alarm CPU and Touch Pad Free VIP Vacuum Insulation Panel VIP Quiet Reliable Compressor Cascade Cooling System vice ic tem able y Energy SAVE Air Filter VIP Plus Vacuum Insulation Panel VIP Filterless Design Insulated Inner Doors ICD Digital Display Remote Alarm CPU and T...

Page 3: ... temperature fluctuations or uneven storage temperatures at different positions within the interior chamber Enhanced temperature uniformity top to bottom front to back assures sample stability at all inventory locations BSL 3 4 or Highly Secured Labs Restricted access to the contained laboratory limits serviceability Twin Guard Series extends critical time necessary to react in the event of mechan...

Page 4: ...on Cabinet design attributes include high strength lockable door latches and doors latchable inner doors adjustable shelves locking casters and seismic restraints to simplify operation installation and to satisfy local codes Compliant to International Standards Assures quality standards safety and performance criteria are met or exceeded Essential for compliance with LEEDs RoHS OSHA other third pa...

Page 5: ...nally lower HVAC loads SANYO is conscious of the need to protect our environment and conserve energy As a corporate pioneer in life science laboratory equipment and appliances and as a global source of solutions ranging from energy manage ment to solar power and alternative energies SANYO remains committed to providing the best possible labora tory equipment for research and clinical needs Uniform...

Page 6: ...p to bottom uniformity Maintains most repeatable cycling wave form for the strictest of GMP applications B 86 C A System A ON Back Up for System B Maintains minimum 65 C reserve temperature B 65 C A System B ON Back Up for System A Maintains minimum 65 C reserve temperature B 65 C A The SANYO Twin Guard Series Freezer can be set for Normal or EcoMode operation EcoMode establishes an overlapping cy...

Page 7: ...l operation and Twin Guard Series backup Reduced Power Consumption The SANYO Twin Guard Series Freezer can be set for Normal or EcoMode operation depending on ambient tem perature and load EcoMode is recom mended for 90 95 of applications Although both refrigeration systems are completely independent EcoMode establishes an overlapping cycle to significantly reduce energy consump tion while optimiz...

Page 8: ...t puts and performance at a glance Multi point data logging offers a graphical display for temperature verification at a glance Precise logic systems control and adjust to temperature setpoints and alarm parameters Setpoint alarm and programmable inputs are established through pop up menus and function keys Logged parameters can be exported to remote databases off site alarm or data capture system...

Page 9: ...failure Solid Tone No Auto Return Touch key is not pressed for 90 minutes Reverts to chamber temperature display None No Door Alarm Door open Notification of error on graphical display Solid tone sounds after 2 minutes Yes New extended range is important for GxP applications to reduce potential incident event reporting The integrated microprocessor control aggregates information from the entire sy...

Page 10: ...ty Inner door latches are standard 7 New air cooled Super Condenser energy transfer technology maintains optimum condenser air flow and eliminates the need for an air filter or periodic cleaning and maintenance required by other manufacturers 8 High impact recessed casters and leveling feet simplify installation 9 Seismic restraints secure cabinet to building superstructure to meet local codes 10 ...

Page 11: ...YO ergonomic design features a high security Easy In Easy Out door latch designed for simplicity and safety Also allows for one handed operation 2 Barrier Film VIP Plus Interior Rigid foamed polyurethane 1 12 4 A vacuum relief valve is mounted in the left wall 17 16 ...

Page 12: ...ion when required RoHS Compliance In 2006 RoHS Restric tion of Hazardous Sub stances legislation EU Directive 2003 95 EU became effective RoHS relates to the restriction of hazardous substances and reductions in environmental pollution Through RoHS legislation the EU and other participating countries are banning toxic substances in electrical equipment such as lead cadmium mercury chro mium 6 PBB ...

Page 13: ...ounts on the cabinet door and is programmable to capture and store non corruptible data points for download ing to an independent data management system to meet criteria outlined in U S FDA21CFR Part II directives An inde pendent probe is placed in the interior chamber Inventory Rack Systems SANYO freezers are available with a variety of inventory racks to meet specific applications Freezers can b...

Page 14: ...3 locations Vacuum Release Port Standard Manual Refrigeration System Refrigeration Technique SANYO 86 Dual Cool independent redundant cooling systems with automatic switchover Compressors Per system Single hermetically sealed 1100W SANYO Cool Safe compressor air cooled Refrigerant Environmentally safe HFC refrigerant Air Exchange Dual condenser fans with particulate trap no filters are required Cy...

Page 15: ...15 Dimensional Drawings MDF U500VXC MDF U700VXC 34 3 870 21 0 533 ...

Page 16: ... installation qualification operational qualification performance qualification calibration and temperature mapping SANYO Connect SANYO s customer driven biomedi cal service program guarantees local attention from qualified SANYO service representatives whenever and wher ever you need it New Unit Installation and Training Preventative Maintenance Warranty and Non Warranty Repairs Calibration Valid...