- 8 -
Connections on PCB
The following shows the connections of connectors on the Temp. controller PCB.
Connector Connects
to
Usage
CN1
Switching power supply
To supply the power to PCB.
CN2
Communication kit
To connect to MTR-480/MTR-L03(option).
CN3
Remote alarm terminal
#1: COM.
#2: N.O.
#3: N.C.
Remote alarm contact outputs.
In normal condition, open between #1 and #2
and closed between #1 and #3.
CN4
#1-#2
㸸
Temp. control relay
#3-#4
㸸
Heater relay
To control chamber temperature (12VDC)
To supply the power to Cap. tube heater
(12VDC)
CN5
#1-#5: Switch PCB
#6-#7: Buzzer
To connect to each switch
To connect to buzzer
CN6
Display PCB
To connect to each LED
CN7
#1-#2: Door switch
#5-#6: AT sensor
#7-#8: Filter sensor
#9-#10: Cascade sensor
To control door open and close
To detect ambient temperature
To detect temperature in condenser outlet pipe.
To detect temperature in cascade.
CN8
#1-#2: Battery
㸦
#1:6V
#2:Battery switch
㸧
#3-#4: Transformer
To supply the power during power failure
To supply the power to PCB.
CN9
#1-#2: H. Comp. relay
To control compressor H ON/OFF (12VDC)
CN10 Unused
CN11
#1-#3: Temp. sensor
To detect chamber temperature.
Summary of Contents for MDF-U55V
Page 14: ...Wiring diagram 11 ...