- 19 -
F00:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in “F00” display to return to chamber temperature
display.
F01:
<Purpose> Setting of high temperature alarm
<Operation> Input F01 and press SET key to display “015” (Factory default).
Setting range is ‘005~020’.
Change a value by pressing key.
Press SET key to memorize the value and to return to chamber
temperature display.
F02:
<Purpose> Setting of low temperature alarm
<Operation> Input F02 and press SET key to display “-10” (Factory default).
Setting range is “-05”~“-20”.
Change a value by pressing key.
Press SET key to memorize the value and to return to chamber
temperature display.
F03:
<Purpose> Display of battery accumulation time
<Operation> Input F03 and press SET key to display alternately F03 with “00.0” (in
case battery used for 36days or less).
Press SET key to return to chamber temperature display.
F04:
<Purpose> Setting of door alarm delay time
<Operation> Press SET key in “F04” to display ‘002’ (Factory default).
setting range is ‘001’~’015’.
Change a value by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
F05:
<Purpose> Compressor turns on with delay when a power is supplied or a power
returns from a power failure.
<Operation> Input “F05” and press SET key to display ‘003’ (Factory default).
Setting range is ‘003’~’015’. (Unit: Minute)
Change a value by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
F06:
<Purpose> Input of service code and reset of accumulation time
<Operation> Input F06 and press SET key to display ‘000’ (Factory default).
Set service code to “384” by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
<Reset of battery accumulation time>
Input service code ‘384’ in F06.
Input ‘409’ to reset battery accmulation time and to return to chamber
temperature display. (Service code is cancelled)
<Reset of condensing fan motor accumulation time>
Input service code ‘384’ in F06.
Input ‘410’ to reset fan motor accumulation time and to return to
chamber temperature display. (Service code is cancelled)
<Cancel> Input F06 again and press SET key to display ‘384’.
Change to ‘000’ by pressing key and key.
Press SET key to memorize the value and to return to chamber
temperature display.
Turn a power off then on to change a value to ‘000’, but it is not stored
in non-volatile memory.
Note) Service code ‘384’ is stored in non-volatile memory during battery
back-up.