1
In [RATIO], select the zoom magnification power.
×1, ×1.4, ×2, ×2.8, ×4, ×5.6, ×8, ×10
2
In [ZOOM TIME], select how long you want the target to be zoomed in.
Select the value within the duration configured in [DURATION] (Step
4
below).
5SEC, 10SEC, 15SEC, 20SEC, 30SEC, 45SEC, 1MIN, 2MIN, 3MIN, 4MIN, 5MIN
The duration configured in [ZOOM TIME] is the period in which zoomed state is retained after the
camera has zoomed in or out to the specified magnification.
After the duration set in [ZOOM TIME] has passed, the zoom magnification before the alarm will be
restored.
Depending on the value selected in [DURATION], the setting in [ZOOM TIME] may be overridden by
the setting in [DURATION].
In [RETURN], select whether to retain the monitoring conditions loaded when an alarm signal
was input even after the alarm action, or return to the previous monitoring conditions.
OFF:
Retains the monitoring condition after the alarm.
ON:
Returns to the monitoring condition before the alarm (auto-recovery function)
If you select “ON” for [RETURN], the system will return to the previous operating conditions before an
alarm action took place, when the alarm duration (DURATION) expires.
When you perform zoom or preset position monitoring operations during the alarm condition, the auto
-recovery function will be disabled.
4
In [DURATION], select how long you want the alarm state to be retained when the
terminal receives an alarm signal and click
SET
.
The terminal will not accept subsequent alarm signals until the set duration expires.
5SEC, 10SEC, 15SEC, 20SEC, 30SEC, 45SEC, 1MIN, 2MIN, 3MIN, 4MIN, 5MIN
CC (Retains the alarm state as long as the alarm signal persists.)
Working with Administrator Configuration Screens 41/71