20
Operating Instructions MD8 Airport
Cleaning and Maintenance
9 Cleaning and Maintenance
9.1 Cleaning
Requirements
The operation has been completed .
WARNING
Moisture on live components can cause electric shock!
Allowing liquids to penetrate inside the housing can cause serious injury and damage
to the device .
t
Do
not
immerse the device in liquid .
t
Prevent liquid droplets from forming during cleaning .
Procedure
t
Disconnect the device from the power supply .
t
CAUTION
There is a risk of injury from contact with substances that are
harmful to health! If the device has come into contact with substances that are
harmful to health: Contact Sartorius Service .
t
NOTICE
Corrosive gases can damage the device . If the device has come into
contact with corrosive gases: Contact Sartorius Service .
t
NOTICE
Corrosion or damage to the device due to unsuitable cleaning agents!
t
Do
not
use corrosive, chloride-containing, and aggressive cleaning agents .
t
Do
not
use cleaning agents that contain abrasive ingredients, e .g . scouring
agents, steel wool .
t
Do
not
use solvent-based cleaning agents .
t
Wipe off the housing with a slightly damp cloth . For more severe contamination,
use a mild soap solution .
9.2 Disinfecting
Requirements
The operation has been completed .
WARNING
Moisture on live components can cause electric shock!
Allowing liquids to penetrate inside the housing can cause serious injury and damage
to the device .
t
Do
not
autoclave the device .
t
Do
not
immerse the device in liquid .
t
Do
not
spray the device with disinfectant; simply wipe clean .
t
Prevent liquid droplets from forming during disinfection .