Problem ...
Causes ...
Solution
The special
– No keys have been
– Press any key
code “
b
“displayed pressed since the balance
does not go out
was turned on
The weight
– Unstable ambient
– Set up the balance
readout changes
conditions
in another area
constantly
– Too much vibration or
– Access the menu to
the balance is exposed
select the appropriate
to a draft
code to adapte the balance
to the particular weighing
environment (see Part 2)
– A door of the draft
– Make sure that all
shield chamber is not
draft shield doors
completely closed
are closed
(only applies to draft shields
with sliding doors)
– The round glass draft
– Position the round
shield is not in place
draft shield
– The draft shield base
– To mount these components,
plate
(4)
or the cen-
see “Getting Started”
tering disk
(27)
is
on page 1–9
not mounted correctly
– The dust cover for
– Press down on the adhesive
balances with a rectangular
strips to attach the dust cover
pan is not attached correctly
and is touching
the weighing pan
– A foreign object
– Remove foreign object
is caught between the
pan and the housing of
the balance
– The cover plate for the
– Turn the cover plate to
port for the below-balance
close the port for
weighing hanger
below-balance weighing
has not been closed
– The sample does not
have a stable weight
(absorbs moisture or
evaporates)
1–29