Weighing in Percent %
This application program enables you to obtain weight readouts in percent, which
are based on a reference weight.
The reference weight readout is stored as a menu-defined percentage (factory
setting: 100%). To change this percentage, see the description on page 4–8.
(The reference percentage will be displayed as long as the reference sample
weight has not yet been stored.)
Determination of the Residual Weight in Percent
Code 2 1 5
Menu codes used in the example:
Function
Code
Weighing in percent
2 1 5
With data ID code
7 2 2
Reference % and weight
7 1 2
Application: quick determination of the dry weight of a sample
Step/(key ID)/key
Readout
Data output/printout
Place container on the pan
+
22.65 g
c
,
t
0.00 g
Add prepared sample
to the container
+
4.61 g
Store ref. wt. as 100%:
+
100.0 %
pRef
+
100 %
(STO REF)
F
Wxx%
+
4.61 g
For moisture analysis,
press
t
at this point
Remove container with
sample; dry sample
xx.x %
Reweigh sample in
the container
+
72.5 %
p
+
72.5 %
Prc
+
72.5 g
Residual weight readout:
w
+
3.34 g
p
+
3.34 g
N
+
3.34 g
The data ID codes
mean:
pRef
+
100 %
reference percentage
Wxx%
+
4.61 g
net reference weight xx%
Prc
+
72.5 g
calculated percentage
N
+
3.34 g
net residual weight
As an alternative, you can select this
output parameter:
Reference weight
7 1 3
4–6