Checking Variations in Weight
Code 2 2 4
– Readout in Percent –
and Code 2 1 5
To store a readout – displayed as a weight or a percent – as the target value,
press
f
. The target value is indicated in the analog display as a bar graph. The
last segment of this bar graph ends exactly below the equal sign of the
checkweighing indication. If the weight of a product is “acceptable,” the last
segment of the bar graph will be within the tolerance range of the checkweighing
indication. You can also read the difference in weight in digits or as a percent.
Menu codes used in the example:
Function
Code
Checking variations in weight 2 2 4
Weighing in percent
2 1 5
With data ID code
7 2 2
Tol target weight
7 1 2
Reference % + ref. weight
Automatic output
4 2 1
Application: checkweighing finished products with automatic output of the
variations within the tolerance limits
Step/(key ID)/key
Readout
Data output/printout
c
,
t
0.00 g
Load part that has the
“target weight”
+
235.09 g
Press (STO REF)
F
+
100.0 %
pRef
+
100 %
Wxx%
+ 235.09 g
To store the limit for checking
tolerances: press (STO)
f
0.0 % Net
Lim
+
1.0 %
Setp
+
235.09 g
Remove part with
ideal target weight
Checkweigh parts
Checkweighing indication
Example:
+
0.7 % Net
Prc1
+
0.7 %
The data ID codes
mean:
Lim
+
1.0 %
over/under limits in percent
Setp
+
235.09 g
target weight stored
pRef
+
100 %
reference percentage
Wxx%
+
235.09 g
reference weight
Prc1
+
0.7 %
calculated variation in percent
As an alternative, you can select
this output parameter:
Target wt. and ref. wt. only
7 1 3
4–11