Counting with Additional Requirements
Changing the Reference Sample Quantity
You can change the reference sample quantity as required. The reference sample
quantity will be displayed above the unit symbol after you press
c
.
To select the change function:
Hold down
i
for more than 2 seconds
To change the setting:
Use
f
identified by – ^ –
If you enter a wrong number, press
c
to clear it.
To store the ref. quantity:
– temporarily until you turn off the power: briefly
press
F
– p
ermanently after you turn off the power (in the non-
volatile memory): press
F
for more than 2 seconds
Note: This setting is not canceled by the reset code 9 – –1
o
.
Changing the reference sample quantity
Code
No change
2 3 1
In cycles: 5,10, 20, 50,100, 5,10, …
* 2 3 2
In increments of one
2 3 3
Storage Parameter for the Reference Sample Weight
The weight is stored
Code
with full accuracy according to the internal resolution
3 5 1
according to the display accuracy
* 3 5 2
Reference Sample Updating
While counting (piece count readout), you can press
F
to update the reference
sample for optimizing the accuracy (identified by – OPT – in the display next to the
key). In the process, the reference weight is recalculated based on the total
reference sample weight and the piece count displayed. It makes sense to use the
reference sample updating function when the weight of a reference sample quantity
is lighter than 100 digits. In this case, you should about double the reference
sample quantity. You can repeat reference sample updating several times.
* = factory setting
4–14