Example:
Determining the density of a solid sample.
Parameter settings:
Applic.-Density-Dec.Plcs-1 Dec. Pl
(menu code 2. 9. 1. 2)
Step
Key (or instruction)
Display/Data output
1. Attach sample holder to suspension wire
2. Tare the balance
(TARE)
0.0
g
3. Start application program
(ENTER)
4. Confirm “
Air
” display
(ENTER)
air
?
5. Determine the weight
of the sample in air:
Place sample on the balance
+
20.0
g
*
6. Store value for weight in air
(ENTER)
7. Remove sample from the balance
water
?
8. Determine weight in liquid:
place sample in holder
9. Confirm “
water
” display
(ENTER)
0.0
g
*
10. Immerse sample in liquid
+
15.0
g
*
11. Store value for weight in
(ENTER)
liquid, view result, and print
+
4.0
o
*
Wa
+ 20.0 g
Wfl
+ 15.0 g
Rho
4.0 o
12. Delete result
(CLEAR)
13. Repeat as desired, starting from Step 3.
39