13
Example
Determine an unknown percentage; store the weight on the balance/scale as a reference
percentage
Settings (changes in the factory settings required for this example):
Menu: Application program: Weighing in percent (menu code
2. 1. 5
)
Menu: Reference percentage 100 % (menu code
3. 3. 5
)
Step
Key (or instruction) Display/Printout
1
1. Turn on the balance/scale
e
1
2. Tare the balance/scale
t
0.0
g
1
3. Display the reference
percentage
v
> 2 sec.
ref 100
1
4. Place the reference weight
+
222.5
g
for 100% on the balance/scale
(here: 222.5 g)
1
5. Start application; if the print
v
+
100.00
%
format is set to include data
ID codes,
the following is printed:
Wxx% + 222.500 g
1
6. Place an unknown weight
+
144.94
%
on the balance/scale
(here: 322.5 g)
1
7. Display weight
v
+
322.5
g
1
8. Display percentage
v
+
144.94
%
1
9. Unload the balance/scale
0.00
%
10. Delete the reference percentage
c
11. Repeat the procedure starting
from step 6, if desired.
Summary of Contents for Basic lite