27
Basic
lite
und Gem
lite
Serie
Model
BL3100, BL1500, BL600, BL12
BL6
BL3
GM3101 GM1501 GM601
Weighing capacity
g
3100
1500
610
12000
6000
3000
Readability
g
0.1
0.1
0.1
1
1
1
Tare range (subtractive)
g
3100
1500
610
12
6
3
Repeatability
<
±g 0.1
0.1
0.1
1
1
1
Linearity
<
±g 0.2
0.2
0.2
2
2
2
Operating temperature range
+10…+30°C (50°F…86°F)
Allowable ambient
operating temperature
+5…+40°C
Sensitivity drift within
+10…+30°C
<
±/K 1·10
–5
2·10
–5
5·10
–5
2.5·10
–5
5·10
–5
10·10
–5
Response time (average)
s
2
1.5
1.5
1.5
1.5
1.5
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update (depends on
the filter level selected)
s
0.2–0.8
External calibration weight
kg
1 (F2)
1 (M1)
0.5 (M1) 5 (M1)
5 (M1)
1 (M1)
(of at least accuracy class…)
lb*
2
2
1
10
10
2
Net weight, approx.
kg/lb1.4/3.1
Pan size
mm
174x143 (6.9x5.6 inches)
Dimensions (WxDxH)
mm
189x251x70 (7.4x9.9x2.8 inches)
AC power source/power
requirements
AC adapter, 230 V or 115 V, +15%…– 20%
Frequency
48–60 Hz
Power requirements,
direct current
V
10 to 20
Power consumption (average)
W
0.75
Approx. hours of operation with:
– 9-volt battery, approx.
h
11
– fully charged
rechargeable battery
h
2.5
Hours of operation with the
YRB08Z rechargeable
battery pack
h
25
*
= only for BL models