10
Application Programs
Counting
Purpose
W ith the Counting program you can determine
the number of parts that each have approximately
equal weight.
Available Features
Store the current weight value to have it loaded
as the preset reference sample quantity next time
you initialize the Counting application
The reference sample quantity can be changed
in the operating menu:
See Configuring the Balance/ Scale
The average piece weight is automatically
output via the optional data interface port after
initialization, if the menu code for Printout
with data ID codes is set
Press
v
to toggle the display between piece
count and weight
Factory Settings
Reference sample quantity: 10 (menu code
3. 3. 2
)
Preparation
Configure the Counting application
in the operating menu:
See Configuring the Balance/ Scale
Menu code
2. 1. 4
Counting
Reference sample quantity:
Menu code
3. 3. 1
5 pcs
Menu code
3. 3. 2
10 pcs
Menu code
3. 3. 3
20 pcs
Menu code
3. 3. 4
50 pcs
Menu code
3. 3. 5
100 pcs
See also Configuring the Balance/ Scale