Setting the SBI Data Output
Data output settings can be made in the Setup menu under “
Data
communications:SBI:Data output
.“
The following options are available:
– The displayed value, with or without stability check
– Automatic output of the displayed value, either with or without stability check,
or automatically at defined intervals
– Output of a printout as configured in the “
Device parameters:
Config. printout:Printer 1
“ or “
Printer 2
“ (see next
page).
You can define the printout content by specifying which blocks of information are
to be included (see “Configuring Printouts”).
In general, the current display value is output (weight with unit, calculated value,
alphanumeric display) unless you configure a user-definable data record.
Generally, data is output only after the weighing instrument has stabilized (factory
setting). Here you can define whether data is output on request or automatically,
at stability or without stability check; and configure a user-defined printout.
The data output for “without stability“ must be set accordingly (SBI: Data output).
If you select time-dependent automatic printout, you need to define the print
interval (in display updates) as well.
Each line of a printout can contain up to 20 characters. The first 6 characters,
called the “data header”, identify the subsequent value. Only 14 characters are
available if you disable the header. This is carried out in the “Line format“ menu
item (see also “Setup Overview“ in “Configuration“).
Automatic Data Output (SBI)
You can have the weight readout printed automatically
1
). This printout can be
generated after a certain number of display updates
2
). You can also configure
whether or not the auto-print function is dependent on the stability of the scale
3
).
The display update frequency depends on both the scale model and the operating
status.
Examples:
N + 153.00 g
Net weight
Stat
Display blank
Stat L
Display underload
Stat H
Display overload
„Data output“ setting:
1) 3) Automatic, without stability or automatic with stability.
Factory setting: Manual after stability; i.e., automatic data output function off.
2)
Time-dependent automatic data output:
Interval: 1, 2, 10 or 100 display updates
Factory setting: 1 display update
100
Operating Instructions Combics Complete Scales
Data Interfaces
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......