GMP-compliant printouts
“Linearization“ printout
--------------------
14.07.2010 13:00
Type CAIS2
Ser.no. 12345678
Vers. 1.02.101110
BVers. 01-63-01
--------------------
Linearization
Wt.1 + 7.00 kg
Wt.2 + 15.00 kg
Wt.3 + 22.00 kg
Wt.4 + 30.00 kg
completed
--------------------
14.07.2010 13:02
Name:
--------------------
Calibration/adjustment printout
--------------------
14.07.2010 13:50
Type CAIS2
Ser.no. 12345678
Vers. 1.02.101110
BVers. 01-63-01
--------------------
External calibration
Targ. + 30.00 kg
Diff. - 0.03 kg
External adjustment
Diff. + 0.00 kg
--------------------
14.07.2010 13:52
Name:
--------------------
Setting the preload printout
--------------------
14.07.2010 13:50
Type CAIS2
Ser.no. 12345678
Vers. 1.02.101110
BVers. 01-63-01
--------------------
Set preload
completed
--------------------
14.07.2010 13:52
Name:
--------------------