If gravity is being entered instead of altitude and latitude, then “
Gravity
“ is
displayed for a brief time after “
CAL
.“
The entered value appears in m/s
2
, here for the “Germany (Zone D)“ setting.
t
Press
J
to confi rm the displayed value or press
(
to cancel the adjustment.
y
You are prompted to place the required weight on the platform (e.g.: 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described
in the chapter entitled “Operation” under “Calibration and Adjustment.“
t
Slide the menu access switch to the right (= “closed“ position) and reattach the
cap.
y
The display goes out and the device restarts. Then weighing mode is active.
If adjustment is carried out using a verifi able confi guration data record, the lines
for display of metrological data (lines 1 and 2) show the data valid for use in legal
metrology, if the menu access switch is closed. See also the chapter “Operation“,
“Confi guration for Use in Legal Metrology.“
36
Operating Instructions Combics Complete Scales
Confi guring Weighing Platforms
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......