t
Press
e
to turn off the device.
t
Press
e
turn the device back on.
y
The Sartorius logo is displayed briefly, after which the device is in normal
weighing mode.
t
Press
(
to unload and zero the scale.
t
Press
J
to start the external adjustment.
C.EXT.U
C.EXT.USE.
y
The display “
C.EXT.U
“
appears briefly.
Note
If the display of geographical data (elevation and latitude or gravity) is activated,
this data is displayed and confirmed each with a press of the
J
key (you can
cancel the calibration/adjustment process using the
(
key). See also “Entering
Geographical Data for Use in Legal Metrology“ in this chapter.
Procedure
t
Press
J
to confirm the displayed value or press
(
to cancel
the adjustment.
y
The target value of the required adjustment weight (6.000 kg in the example) is
displayed as a negative value.
t
Place the required adjustment weight on the platform.
If the calibration/adjustment sequence is set to “Cal. then auto adjust”, refer to
the note under “External Calibration/Adjustment with Factory-Defined Weight
(Default Weight).”
y
After a brief pause, the difference since the last span adjustment is displayed
(calibration).
t
If you only want to perform a calibration, press the
(
key to cancel the
calibration/adjustment procedure.
t
Press
J
to adjust the scale.
y
At the conclusion of the calibration procedure, the calibration weight is dis-
played as a positive value.
--------------------
14.01.2010
13:50
Type
CAW3P1-6DC-LCE
Ser.no.
12345678
Vers.no.
1.02.101110
BVers. 01-63-02
--------------------
External calibration
Nom. + 6.000 kg
Diff. +
0.010 kg
External
adjustment
Diff. + 0.000 kg
--------------------
14.01.2010 13:52
Name:
--------------------
When calibration/adjustment is complete, the GMP-compliant printout shown on
the left is generated. If the adjustment procedure is canceled (only calibration is
performed), the last two lines, “External calibration” and “Diff. + 0.000 kg” are not
printed.
t
Unload the scale.
t
Press
e
to turn off the device.
t
Press
e
turn the device back on.
y
The Sartorius logo is displayed briefly, after which the device is in normal
weighing mode.
42
Operating Instructions Combics Complete Scales
Configuring Weighing Platforms
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......