Procedure
t
Remove the cap.
t
Slide the menu access switch to the left (= “open“ position).
t
Activate the Service mode, see “Service Mode.“
t
Select weighing platform “
WP 1
“ in the “
Device Parameters
“
menu item.
t
If the “
Internal
“ setting is not already activated (marked by “
o
“), select
the setting using the “
Q
“ or “
q
“ soft key and confirm with “
O
.“
y
The menu for the “
WP-1 INTERNAL
“ device parameters is displayed.
t
Use the “
Q
“ or “
q
“ soft key to select and open the “
Calibration/
adjustment
“ menu using “
O
.“
t
Use the “
Q
“ or “
q
“ soft key to select and use “
O
“ to open the “
Adjust
without weights
“ menu.
t
Use the “
O
“ soft key to open the “
Input parameters
“ menu.
y
The Input menu is displayed.
Enter the nominal capacity and resolution of the load cells in kg and the
sensitivity of the load cells in mV/V in the corresponding input fields.
The maximum capacity is usually less than the value to be entered in the
“
Adjust without weights
“ menu for the nominal capacity of
the load cell, as load cells carry additional weight (e.g., a weighing platform).
44
Operating Instructions Combics Complete Scales
Configuring Weighing Platforms
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......