If the weighing platform has multiple load cells, multiply the nominal capacity
accordingly.
Example:
The weighing platform has 4 load cells each at 50 kg.
The nominal capacity is 4 x 50 kg = 200 kg.
In the example shown here, the weighing platform consists of one load cell with a
maximum capacity of 10 kg.
t
Press
10.00
and confirm with the “
l
“ soft key.
y
The “
Resolution
“ input field is selected.
The smallest digit “d“ is entered in this field in “kg“. The value must correspond
to the “
D
“ entry in “
ADC configuration: Standard:
Ranges: Single-range mode
.“
In the example, this should be d = 0.002 kg.
t
Press
0.002
and confirm with the “
l
“ soft key.
y
The “
Sensitivity 1
“ input field is selected.
If the weighing platform has multiple load cells, enter the sensitivity either
– under “
Sensitivity 1
“ … “
Sensitivity 4
“ as an individual
value or
– under “
Sensitivity 1
“ as an average value.
If an average value is entered for all cells, or if fewer than 4 load cells are
present, enter “0“ in the remaining fields. Range of permitted values: 0.01…5
mV/V.
t
Press
1.944
, e.g. to enter a sensitivity of 1.944 mV/V.
t
Confirm using the “
l
“ soft key.
y
The “
Sensitivity 2
“ input field is selected.
t
Either enter a value or set all remaining input fields to “0.“
t
Confirm each using the “
l
“ soft key.
t
Press the “
o
“ soft key to exit the menu.
t
Use the “
q
“ soft key to select the “
Save parameters
“ submenu.
t
To save the configuration, select “
Yes
“ and confirm using the
l
soft key.
In the first line of the display, the message “
Data stored
“ is shown
briefly.
Then the program then returns to the “No“ display status.
t
Use the “
o
“ to return to the next menu level.
t
Press
M
or “
oo
“ to exit the Setup menu.
t
Slide the menu access switch to the right (= “closed“ position) and reattach the
cap.
y
The display goes out and the device restarts. Then weighing mode is active.
Operating Instructions Combics Complete Scales
45
Configuring Weighing Platforms
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......