Function Allocation of the
J
Key for Linearization and Setting/
Deleting the Preload
Purpose
The
J
key is normally used for the calibration/adjustment function.
The following additional functions can be allocated to the key when the Service
mode is activated:
– External linearization with default weights
– External linearization with entered linearization weights
– Internal linearization (for external IS platforms only)
– Set preload
– Delete preload
3
Once linearization has been completed, or after a preload has been set or deleted
the function of the
J
key must be reallocated back to its original function in the
Setup menu, e.g. external calibration/adjustment with default weights.
Procedure
t
Remove the cap.
t
Slide the menu access switch to the left (= “open“ position).
t
Activate the Service mode, see “Service Mode.“
t
Select weighing platform “
WP 1
“ in the “
Device Parameters
“
menu item.
t
If the “
Internal
“ setting is not already activated (marked by “
o
“), select
the setting using the “
Q
“ or “
q
“ soft key and confirm with “
O
.“
y
The menu for the “
WP-1 INTERNAL
“ device parameters is displayed.
t
Use the “
Q
“ or “
q
“ soft key to select and open the “
Calibration/
adjustment
“ menu using “
O
.“
y
The “
CAL key function
“ submenu is displayed.
t
Use the “
Q
“ or “
q
“ soft key to select the corresponding menu item and confirm
with “
l
.“
y
The menu item is marked by a circle “
o
.“
Note
The functions that can be configured in the “
CAL key function
“
submenu depends on the selected weighing platform and its configuration data.
Functions that cannot be accessed are not displayed in the selection screen.
Note
The function set in the “
CAL key function
“ menu is carried out in the
normal weighing mode because when you exit the Setup menu the Service mode is
deactivated. To perform the function on a digital weighing platform (such as an IS
platform), however, it must be carried out in Service mode.
t
The procedure is as follows: after selecting the desired function and exiting
the Setup menu, reactivate Service mode again and then exit the Setup menu
immediately by pressing the
M
key or the “
oo
“ soft key.
y
The scale is now in Service mode without this being displayed.
t
Trigger the previously set function using the
J
key.
y
The display shows “
S_CAl:
“ indicating that the scale is in Service mode.
t
The corresponding menu item can be selected using the “
Select
“ soft key
and carried out using the
J
key.
t
If you cancel the function using the
(
key or by restarting the scale using
the
e
key, you will exit the Service mode.
46
Operating Instructions Combics Complete Scales
Configuring Weighing Platforms
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......