External calibration and manual adjustment with default weights
(weighing parameters: factory settings)
Preparation
t
Select Setup menu: Press the
M
key.
t
Select Device parameters: Press the “
O
“ soft key.
t
Select weighing platform “
WP 1
,“ “
WP 2
:“ Press the “
O
“ soft key or
t
Select interface “
COM 1
,“ “
COM 2
“ or “
UNICOM
“ (depending on the
interface): Press the “
O
“ soft key.
t
Select weighing platform “
WP 3
:“ Press the “
O
“ soft key.
Calibration/Adjustment
CAL key function
o Ext. cal./adj.; factory-def. wt.
Ext. cal./adj.; user-defined wt.
Key blocked
Cal./adj. sequence
Cal. then auto adj.
o Cal. then manual adj.
isoCAL function (for Sartorius IS weighing platforms only)
o Off
Adjustment prompt
Activate ext. adj.
o Activated
Deactivated
Parameter for external weight
o = Factory setting
t
To save settings and exit the Setup menu: Press the
M
key or the “
oo
“
soft key.
Procedure
t
Zeroing the scale by pressing the
(
key.
t
Start calibration/adjustment using the
J
key.
C.EXT.D
C.EXT.DEF.
12
y
The display “
C.EXT.D
“ appears for two seconds.
y
You are prompted to place the required calibration/adjustment weight on the
platform (e.g., 10 ,000 g).
88
Operating Instructions Combics Complete Scales
Operation
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......