t
Place the calibration/adjustment weight on the weighing platform.
y
The difference between the measured value and the true weight of the sample
will be displayed with plus/minus signs.
External calibration
Nom. + 10000 g
Diff. + 1 g
y
A printout will be generated if the adjustment is not carried out and the
procedure is stopped by pressing the
(
key.
t
Start adjustment via the
J
key (cancel calibration/adjustment via
(
).
y
The adjustment weight is displayed once adjustment is finished.
--------------------
24.10.2010 10:15
Type CAIS3
Vers.no. 1.02.101110
BVers. 01-63-02
--------------------
External calibration
Nom. + 10000 g
Diff. + 1 g
External adjustment
Diff. + 0 g
--------------------
24.10.2010 10:15
Name:
--------------------
y
A GMP-compliant printout is generated.
Operating Instructions Combics Complete Scales
89
Operation