Installation
CAH3 models: Remove transport locks
t
Place the weighing platform at its installation location, remove the
weighing pan.
t
Remove the transport lock: Remove screw
1
t
Loosen screw
2
.
t
Turn the mounting bracket 180° and re-secure screw
2
.
t
Re-attach screw
1
to the lever
3
The transport lock must be re-installed before transporting the weighing platform.
Leveling the Weighing Platform
The weighing platform must be exactly level to ensure reproducible weighing results
every time. Therefore, the weighing platform must always be re-leveled after it has
been moved to a different location.
t
Remove the weighing pan.
t
Loosen the lock nuts using a wrench (SW17).
t
Use a SW5 Allen key to screw the leveling feet in/out.
Turning the leveling feet clockwise lifts the weighing platform,
turning the leveling feet counterclockwise lowers the weighing platform.
Operating Instructions Combics Complete Scales
9
Summary of Contents for Combics CAH3
Page 130: ...130 Operating Instructions Combics Complete Scales Declarations of Conformity ...
Page 131: ...Operating Instructions Combics Complete Scales 131 Declarations of Conformity ...
Page 132: ...132 Operating Instructions Combics Complete Scales Test Certificate ...
Page 133: ...Operating Instructions Combics Complete Scales 133 Plates and Markings ...
Page 134: ...134 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 135: ...Operating Instructions Combics Complete Scales 135 Plates and Markings ...
Page 136: ...136 Operating Instructions Combics Complete Scales Plates and Markings ...
Page 139: ...Operating Instructions Combics Complete Scales 139 ...
Page 144: ......
Page 146: ......
Page 147: ......