Operating Instructions Cubis
®
MCE
37
System Settings
7.3.2 Parameters in the “Device” Main Menu
Parameters in the “Extras” Submenu
Parameters
Setting values
Explanation
MENU
EDITABLE*
Activates write access: The menu parameters can be changed.
RD.ONLY
Activates read access: The menu parameters
cannot
be changed.
SIGNAL
OFF
Switches the acoustic signal off.
QUIET
Sets the volume of the acoustic signal to “quiet”.
MEDIUM*
Sets the volume of the acoustic signal to “medium”.
LOUD
Sets the volume of the acoustic signal to “loud”.
KEYS
UNLOCKED*
Deactivates the key lock.
LOCKED
Activates the key lock. Only the [On | Off], [Menu], [Zero], and [Tare] keys can
be used.
ON MODE
ON/STDBY*
−
The [On | Off] key switches between on | standby.
− Switch on with the [On | Off] key.
ON/OFF
−
The [On | Off] key switches between on | off.
−
Switch on with the on key.
AUTO ON
Changes the function of the [On | Off] key: The device
no longer
switches off
or to standby mode, instead it starts a boot process.
LIGHTING
OFF
Deactivates the lighting on the operating display.
ON*
Activates the lighting on the operating display.
* Factory setting
Parameters in the “Leveling” Submenu
Parameters
Setting values
Explanation
LEVEL
OFF
The device does not display messages on the status of the level.
NOTE*
The [LEVELING] display flashes if the device needs to be leveled.
ERR.MSG.
The [LEVEL.ERR.] message appears in the operating display if the device has
not been leveled.
START
KEY
Sets the trigger for leveling to “key”: Leveling must be started manually by
pressing the [Leveling] button.
AUTO*
Sets the trigger for leveling to “automatic”: The device automatically performs
a leveling function using the integrated sensor before every internal
calibration | adjustment.
* Factory setting