56
Operating Instructions Cubis
®
MCE
Operation
t
To switch the displayed mass unit during weighing or before starting an
application: Press the “Toggle between weight units” button or the
“Change” key multiple times if necessary, until the desired unit is dis-
played.
y
The current weight value is displayed in the selected unit.
8.11.2 Running the “Statistics” Application
The “Statistics” application saves up to 99 weight values and evaluates these
statistically.
The following values are saved and exported by the statistics application:
−
Number of components
−
Mean value
−
Standard deviation
−
Variation coefficient
−
Sum of all values
− Lowest value (minimum)
− Highest value (maximum)
−
Spread: Difference between maximum and minimum
The “Statistics” application can be combined with the following functions:
−
Toggle between weight units only available before starting the
application
−
ID marking
−
Automatic taring
Procedure
t
Select the Statistics application in the “APPLIC.” menu.
t
Exit the settings menu.
t
Place a sample on the weighing pan.
t
Change the displayed weight unit if necessary.
t
To start the recording of the statistics: Press the [Confirm] button on
the keypad.
y
The current weight value is saved.
t
To save the next value:
t
Remove the sample on the weighing pan.
t
Place a new sample on the weighing pan and press the [Confirm]
button.
t
To switch between the display of the current weight value, the number of
saved components, and the calculated mean in the results display: Press
the [Up] or [Down] button.