Operating Instructions Cubis
®
MCE
41
System Settings
7.3.3 Parameters in the “Data Output” Main Menu
Parameters in the “SBI Communication” Submenu
Parameters
Setting values
Explanations
COM. OUTP.
IND.NO*
Starts the data output after pressing the key or software command as an
individual value without stability.
IND.AFTER
Starts the data output after pressing the key or software command as an
individual value without stability.
AUTO W/O
Activates the automatic data output without stability.
AUTO W/
Activates the automatic data output after stability.
STOP.AUTO
OFF*
Deactivates the option to abort the automatic data output.
ON
The automatic data output is aborted by pressing the [PRINT] key or a
software command.
AUTO.CYCL.
EACH VAL*
Starts the automatic data output with a cycle after each value.
2. VALUE
Starts the automatic data output with a cycle after every 2nd value.
INTERV.
Starts the automatic data output at the output rate set under “INPUT / INTERV.”
FORMAT
16 CHARS
Data output provides 16 characters per line (16 characters for the measured
values only).
22 CHARS*
Data output provides 22 characters per line (16 characters for the measured
values and 6 characters for identifiers).
EXTR.LINE
Data output provides an additional line with the date, time, and weight value.
AUTO.TARE
OFF*
Deactivates automatic taring after data output.
ON
The device automatically tares after data output.
* Factory setting
Parameters in the “Settings for Printouts” Submenu
Parameters
Setting values
Explanation
ACTIVATE
MAN. NO
Manual without stability: Print process can be started manually at any time.
MAN.AFTER*
Manual after stability: After pressing the [Print] key, the print command is
only executed once stability is achieved.
INTERV.
Starts the automatic data output at the output rate set under “INPUT / INTERV.”
AUTO.LC
Automatically at load change: Print process starts after every load change.
FORMAT
22 CHARS*
Printer output prints 22 characters per line (16 characters for the measured
values and 6 characters for identifiers)
EXTR.LINE
Printer output prints an additional line with the date, time, and weight value.
* Factory setting
Summary of Contents for Cubis MCE11201S
Page 2: ......
Page 78: ...78 Operating Instructions Cubis MCE Conformity...