Technical Data
Unit
Value
Relative humidity
At temperatures of up to 31°C
%
80
Then linear decrease from 80% at 31°C to 50% at 40°C
No
heat from heating systems or direct sunlight
No
drafts from open windows, AC systems, or doors
No
vibrations
No
“heavy traffic” areas (personnel)
No
electromagnetic fields
No
dry air
15.3.2 Ambient Temperature for the isoCAL Function
Unit
Value
Scope of application as per Directive 2014/31/EU
With isoCAL function
°C
+10 – +30
Without isoCAL function
°C
+10 – +30
15.3.3 Protection Class
Value
IP protection according to IEC 60529-1
IP 54
15.4 Materials
Housing: Die-cast aluminum and stainless steel 1.4401 | 1.4404
Control unit: Plastic PBT, float glass and stainless steel 1.4301
15.5 Integrated Clock
Unit
Value
Maximum deviation per month (RTC)
s
30
15.6 Backup Battery
Unit
Value
Lithium battery, type CR2032
Service life at room temperature, minimum
Years
10