3. Check the pH/mV meter for correct functioning
as follows:
Connect the shorting cap to the BNC input.
Select mV mode and record the mV value:
Press the
mV
softkey, if necessary.
Make sure the portable meter is in mV mode.
If the mV value is 0 ± 0.3, the meter is functioning
correctly. Note that the long-term drift specified for
this portable meter is 0.1 mV/month since the most
recent calibration.
§
Connect the electrode to the input connector
(1)
on the meter (push in and twist to lock).
For accurate measurements, you can connect an
additional ATC probe to the ATC port
(2)
.
Female connector for Sartorius DocuClip
®
(3)
:
Automatic electrode recognition, automatic
documentation and storage of calibration data.
4. Rinse and blot-dry electrodes between measure-
ments (do not wipe). Rinse electrodes with distilled
water, deionized water, or some of the next test
solution.