Verified Models with EC-type Approval: Specifications
Model
ED6202S-0CE, ED4202S-0CE,
ED3202S-0CE, ED2202S-0CE
GW6202-0CE,
ED4202S-PCE
GW3202-0CE,
GW6202-0CEDS
GW3202-0CEDS
Type
BD ED 200
BD ED 200
BD ED 200
BD ED 200
Accuracy class
1
)
K
K
K
K
Weighing capacity, Max
1
) g
6200
4200
3200
2200
Scale interval d
1
)
g
0.01
0.01
0.01
0.01
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
) g
0.1
0.1
0.1
0.1
Minimum capacity, Min
1
) g
0.5
0.5
0.5
0.5
Response time (average)
s
1.1
1.1
1.1
1.1
Range of use according
to CD
1
)
g
0.5 – 6200
0.5 – 4200
0.5 – 3200
0.5 – 2200
Allowable ambient
°C
+10 to +30 (+50°F to +86°F)
operating temperature
Net weight, approx.
kg
3.5; GW...-0CEDS: 4.9
Weighing pan size
mm
180
+
180; GW…-0CEDS: 150
d
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91; GW…-0CEDS: 230
+
303
+
260
Model
ED822-0CE
ED8201-0CE
ED5201-0CE
ED2201-0CE
Type
BD ED 200
BD ED 200
BD ED 200
BD ED 200
Accuracy class
1
)
K
K
K
K
Weighing capacity, Max
1
) g
820
8200
5200
2200
Scale interval d
1
)
g
0.01
0.1
0.1
0.1
Tare range (subtractive)
<
100% of the maximum capacity
Verification scale interval e
1
) g
0.1
1
1
0.1
Minimum capacity, Min
1
) g
0.5
5
5
5
Response time (average)
s
1.1
1
1
1.1
Range of use according
to CD
1
)
g
0.5 – 820
5 – 8200
5 – 5200
5 – 2200
Allowable ambient
°C
+10 to +30 (+50°F to +86°F)
operating temperature
Net weight, approx.
kg
3.5
Weighing pan size
mm
180
+
180
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91
1
) CD = Council Directive 90/384/EEC for non-automatic weighing instruments;
applicable to the European Economic Area
55