38 Entris
®
II Advanced Line Operating Instructions
Operation
8 Operation
8 .1
Switching the Device On/Off
(Standby)
Requirements
The device is connected to the power supply .
Procedure
t
NOTICE
Pointed or sharp-edged objects may
damage the operating display! Only touch the
operating display with your fingertips .
t
To switch the device on: Press the [On/Off] button .
t
To switch the device to standby mode: Press and
hold the [On/Off] button .
8 .2 Waiting for the Warm-up Time
In order to provide accurate weighing results, the device
must have reached the necessary operating
temperature . The operating temperature is reached
after the warm-up time . The warm-up time starts when
the device is switched on .
Procedure
t
Switch on the device .
t
Wait until the operating temperature is reached (for
warm-up time, see Chapter “14 .8 Warm-up Time”,
page 48) .
The warm-up time must be observed for a
conformity-assessed device, otherwise the
weight values cannot be used .
8 .3 User Login
If user profiles are created: A user login is required every
time the device is switched on .
Procedure
t
Switch on the device .
t
Press the user profile
selection (1) .
t
Select a user profile,
e .g ., Administrator .
t
Press the [Apply] button (2) .
y
If a password is assigned: The input screen for the
password opens .
t
Enter the password and confirm with the [Confirm]
button .
t
To log in a different user: Press the [On/Off] button .
y
The device is in standby mode .
t
Press the [On/Off] button .
y
The device is switched on .
y
User Management is displayed .
t
Log in a new user .
8 .4 Leveling the Device with an
Electronic Level Indicator
Unevenness at the installation site of the device may
result in incorrect weighing results . Leveling the balance
compensates for unevenness at the installation site by
twisting the front leveling feet of the balance .
Procedure
t
Follow the instructions of the Leveling Wizard .
8 .4 .1 Level Indicator Test
The level indicator test is used to check the movement
of the level indicator . The level indicator must be
positioned at three different places during the level
indicator test . The level indicator is positioned by
rotating the front leveling feet .
Procedure
t
Open the main menu .
t
Press the “Leveling” button
t
Press “Level test” .
t
Follow the Level Indicator Wizard’s instructions .
1
2
Summary of Contents for Entris II Advanced BCA124-1S