52 Entris
®
II Advanced Line Operating Instructions
Technical Data
14 .12 .3 Models BCA5201 | BCA2201
BCA5
201-
1x
(x
= only
S, SAR, S
JP
, SKR)
BCA5
201I-
1x
|
BCA5
201-
1x
(x
= only
C
CN, OIN)
BCA2201-
1x
(x
= only
S, SAR, S
JP
, SKR)
BCA2201I-
1x
|
BCA2201-
1x
(x
= only
C
CN, OIN)
Unit
Value
Value
Value
Value
Readability | Scale interval (d)
mg
100
100
100
100
Maximum capacity (Max)
g
5,200
5,200
2,200
2,200
Repeatability
At 5%, typical value
mg
50
50
50
50
At approx . the maximum load, typical value
mg
100
50
100
50
Linearity deviation
Limits
± mg
300
100
300
100
Typical value
± mg
100
60
100
60
Sensitivity drift b10°C and +30°C
±
ppm/K
7
2
7
2
Tare maximum capacity (subtractive)
<100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change
K
2
2
Time interval
h
6
6
Only for models with approval:
Accuracy class
II
II
Type
BC-AE
BC-AE
Verification scale interval (e)
mg
100
100
Minimum load (Min)
mg
5,000
5,000
Minimum initial weight according to USP (United
States Pharmacopeia), Chap . 41
Optimum minimum initial weight
g
82
82
82
82
Typical stabilization time
100
100
100
100
Typical measurement time
s
≤1.0
≤1.0
≤1.0
≤1.0
Typical stabilization time
s
≤0.9
≤0.9
≤0.9
≤0.9
Recommended calibration weight
External calibrated test weight
g
5,000
5,000
2,000
2,000
Accuracy class in accordance with OIML R111-1
F2
F2
F2
F2