50 Entris
®
II Advanced Line Operating Instructions
Technical Data
14 .12 Metrological Data
14 .12 .1 Models BCA6202 | BCA4202 | BCA3202
BCA6202I-
1x
|
BCA6202
-1x
BCA4202I-
1x
|
BCA4202
-1x
BCA3202I-
1x
|
BCA3202I-
1x
Unit
Value
Value
Value
Readability | Scale interval (d)
mg
10
10
10
Maximum capacity (Max)
g
6,200
4,200
3,200
Repeatability
At 5%, typical value
mg
5
5
5
At approx . the maximum load, typical value
mg
10
10
10
Linearity deviation
Limits
± mg
20
20
20
Typical value
± mg
6
6
6
Sensitivity drift b10°C and +30°C
±
ppm/K
2
2
2
Tare maximum capacity (subtractive)
<100% of maximum capacity
isoCAL (only for I-1x models):
Temperature change
K
2
2
2
Time interval
h
6
6
6
Only for models with approval:
Accuracy class
II
II
II
Type
BC-AE
BC-AE
BC-AE
Verification scale interval (e)
mg
100
100
100
Minimum load (Min)
mg
500
500
500
Minimum initial weight according to USP (United
States Pharmacopeia), Chap . 41
Optimum minimum initial weight
g
8 .2
8 .2
8 .2
Typical stabilization time
10
10
10
Typical measurement time
s
≤1.0
≤1.0
≤1.0
Typical stabilization time
s
≤0.9
≤0.9
≤0.9
Recommended calibration weight
External calibrated test weight
g
5,000
2,000
2,000
Accuracy class in accordance with OIML R111-1
F1
F1
F1
Summary of Contents for Entris II Advanced BCA124-1S
Page 324: ......