Technical Note 

Your balance is equipped with a 
serial RS232 interface port for 
connection to a computer or 
other peripheral device.
You can connect a computer to 
change, start and/or monitor the 
functions of the balance and the 
application programs.

Features

 -

Type of interface: Serial interface

 -

Operating mode: Full duplex

 -

Standard: RS-232

 -

Transmission rates: 600, 1200, 2400, 4800, 9600, 19200, 
38400, 57600 and 115200 baud

 -

Parity: odd, even, none

 -

Number of data bits: 7 or 8 bits

 -

Character format: 
1 start bit, 7-bit ASCII, parity, 1 Stop bit

 -

Handshake:  
For 2-wire interface: Software 
For 4-wire interface: Hardware (CTS/RTS)

 -

Data output format of the balance:  
16 or 22 characters

Factory Settings

 

-

Transmission rate

9600

 baud (Code 2.1.2.7.)

 -

Parity: 

ODD

 (Code 2.1.3.3.)

 -

Stop bits: 

1 STOP BIT

 (Code 2.1.4.1.)

 -

Handshake: 

HRDWARE.

 Hardware (Code 2.1.5.2.)

 -

Operating mode: 

SBI

 (Code 2.1.1.1.)

 -

Printing: 

IND.NO. 

Manual without stability (Code 3.1.1.1.)

Preparation

See “Pin Assignments” and “Pin Assignment Chart”

Summary of Contents for Entris II

Page 1: ...1200 2400 4800 9600 19200 38400 57600 and 115200 baud Parity odd even none Number of data bits 7 or 8 bits Character format 1 start bit 7 bit ASCII parity 1 Stop bit Handshake For 2 wire interface Sof...

Page 2: ...9 10 11 12 13 14 15 16 CR LF or H i g h or L o w or C a l E x t Space Cal Ext Calibration external High Overload Low Underload Error Codes Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 E r r CR LF A...

Page 3: ...tified on the printout set the following parameters Communication PRINTER YDP20 0CE YDP30 Code 2 1 1 x Non verified digits are marked by square brackets SBI mode When the SBI Code 2 1 1 1 mode is acti...

Page 4: ...K Ambient conditions very stable L Ambient conditions stable M Ambient conditions unstable N Ambient conditions very unstable O Block keys P Print auto print activate or block Q Beep R Unblock keys S...

Page 5: ...y PRINT or by a software command Esc P or Esc kP Automatic Data Output Activate the auto print Code 3 1 1 4 3 1 1 5 operating code to have data output to the interface port without a print command You...

Page 6: ...connect any other pins to the cable connector of the balance Cable type AWG 24 specification Waage Computer 9 pol 25 pol Stecker Buchse TxD 2 3 RxD 3 2 CTS 7 4 RTS 8 5 GND 5 7 Waage Computer 9 pol 9 p...

Page 7: ...7...

Page 8: ...GmbH Co KG Status 01 2021 Germany Sartorius Lab Instruments GmbH Co KG Otto Brenner Strasse 20 37079 Goettingen Phone 49 551 308 0 USA Sartorius Corporation 565 Johnson Avenue Bohemia NY 11716 Phone 1...

Reviews: