Interface Description
Purpose
Your Factory scale comes equipped with
an interface port for connection to a
computer or other peripheral device.
You can use an on-line computer to
change, start and/or monitor the func-
tions of the scale and the application
programs. The interface port also has
four data output port lines for the
over/under checkweighing program.
A connector for a hand or foot switch
is also provided.
!
Warning When Using Pre-wired
RS-232 Connecting Cables!
RS-232 cables purchased from other
manufacturers often have incorrect
pin assignments for use with Sartorius
scales. Be sure to check the pin assign-
ment against the chart below before
connecting the cable, and disconnect
any lines marked “Internally Connected”
(e.g., pin 6). Failure to do so may
damage or even completely ruin your
scale and/or peripheral device.
Available Features
Type of interface:
Serial interface
Operating mode:
Full duplex
Standard:
RS-232
Transmission rates:
150; 300; 600; 1,200; 2,400; 4,800; 9,600;
19,200 baud
Parity:
Space, odd, even
Character format:
1 start bit, 7- or 8-bit ASCII, parity, 1 or 2 stop bits
Handshake:
2-wire interface: via software (XON/XOFF);
4-wire interface: via hardware handshake lines
(CTS/DTR)
(CTS/DTR)
Operating mode:
SBI, XBPI*
Network address**:
0, 1, 2, ..., 30, 31
Data output format of the scale:
16 or 22 characters
* XBPI operating mode: 9,600 baud, 8 bits, odd parity, 1 stop bit
** Network address is valid only in the XBPI mode
Factory Settings:
Transmission rate:
1,200 baud
Parity:
Odd
Stop bits:
1 stop bit
Handshake:
Hardware
1 character after CTS
Operating mode:
SBI
Network address:
0
Print manually/automatically:
Manual after stability
Stop automatic printing:
Not possible
Automatic printout,
After 1 display update
time-dependent:
Tare after ind. printout:
Off
Application initialization values:
Off
Line format:
For other applications/GLP
(22 characters)
Preparation
§
See the page 113 for the pin assignment chart