Lift-up Mechanism (Option T10)
§
Align the transverse strut (3) with four screws M5
+
8 so that it
is perpendicular to the left side plate (1) and right side plate (2),
and screw the strut and side plates together.
§
Guide the scale measurement cable through the left strut.
Do not separate the scale platform and display device; otherwise
the scales will no longer be calibrated for use in legal metrology.
§
Drive the sleeves (17)
L = 42 mm into the left
and right base plate (8).
The measurement
cable should be located
between the sleeve and
the side plate.
§
Screw the left base plate (7) and the left side plate (1) together,
and the right base plate (8) and right side plate (2) together,
using screw M12 + 65 and nut M12.
§
Place the lift-up mechanism in an upright position on the instal-
lation site and secure the base plates (7 and 8) to the floor with
four heavy-duty dowels (fastening anchors M10
+
90). It must
be checked in advance that the dowels supplied are suitable for
the floor surface.
§
Drive the fastening anchors in completely and screw together
with nut M10.
Important:
Concrete that can withstand the load is required; otherwise the
anchor will not remain stable.
Important:
Cut off the protruding end of the dowel.
8
Hülse
M10
+
55
1
1
2
8
7
3
2