9
Leveling the Weighing Platform Using the Level Indicator
At the place of Installation, you will need to loosen the locknuts or knurled
nuts on the leveling feet using the open-end wrench (Spanner) that is
supplied with the scale. Then level the weighing platform using the leveling
feet (3) so that the air bubble is centered within the circle of the level
indicator (4):
To lift the weighing platform, extend the leveling feet (turn clockwise). To
lower the weighing platform, retract the leveling feet (turn
counterclockwise).
Check to ensure that the 3 leveling feet touch the surface. The leveling feet
must equally support the weight of the platform.
After leveling the weighing platform, retighten the locknuts securely against
the platform frame.
Installing the Weighing Platform in a Pit:
The level indicator can be detached. To find out how, contact your local
Sartorius dealer or service technician.
Before operating the scale, connect the following cables and completely
assemble the components listed below:
—
interface cable to a Sartorius isi industrial terminal or a Computer
with special Software
—
separate power cable for power supply via a Sartorius AC adapter
(optional)
—
components for conveyor Systems
Make absolutely sure to unplug the weighing platform from the
power supply