17
Specifications Chart
Access the configuration menu of the connected suited Sartorius terminal to set the code for the group of specifications of your
choice. You must set one of the appropriate codes before verification of the platform as a legal measuring instrument.
Please note that when you change the configuration, this may also change the accuracy class and the preload range
besides the weighing capacity and the readability.
Then select the appropriate ID label and affix it to the weighing platform.
For directions, refer to the section entitled "Selecting and Affixing the Respective ID Label."
IS FEG-S0CE: (Type HA BD)
Weighing range structure
SuperRange (High-Resolution)
Multi-Interval
SingleRange
Weighing range
Standard
Configurable
Standard
Configurable
Maximum capacity,
Max kg
64 51 31 16
64 64 64 31 16
Scale interval g
(readability)
1
1
1
1
0 … 15kg 5
15 … 30kg 10
30 … 64kg 20
10 20 10 5
Zero tracking ± 2% kg ± 1.28 ± 1.02 ± 0.62 ± 0.32
± 1.28 ± 1.28 ± 1.28 ± 0.62 ± 0.32
Initial zero-setting
range kg
± 1.3
± 1.3
± 1.3
± 1.3
± 1.3
± 1.3
± 1.3
± 1.3
± 1.3
Preload
1
) kg
0 13 33 48
0 0 0 33 48
ISO test function
2
)
yes yes yes yes
yes yes yes yes yes
Metrological specifications:
Accuracy class
– –
Minimum capacity, Min,
acc. to CD
3
) g
50
50
– –
100
200
400
200
100
Verification scale
interval g
10 10 – – 5/10/20 10 20 10 5
Temperature range
o
C
0 … + 40
–
–
0 … + 40
Select particular group of
specifications (use code)
Sp.grp.1*
(10 3 1)
Sp.grp.2
(10 3 2)
Sp.grp.3
(10 3 3)
Sp.grp.4
(10 3 4)
Sp.grp.5
(10 3 5)
Sp.grp.6
(10 3 6)
Sp.grp.7
(10 3 7)
Sp.grp.8
(10 3 8)
Sp.grp.9
(10 3 9)
* = factory setting
1
) No reduction in the maximum capacity up to this preload / higher preloads reduce the maximum capacity
in 1-kg steps (you cannot set a preload > 49 kg)
2
) The ISO lest function cannot be selected for preloads S 32 kg
*) CD = Council Directive 2009/23/EC on non-automatic weighing instruments used in the EU and the Signatories of the
Agreement on the European Economic Area
4) Not allowed for use as a legal measuring instrument (legal for trade)
A 50-kg calibration weight is required for external adjustment