Connecting Electronic Peripheral Devices
●
Make sure to unplug the balance from AC power
before you connect or disconnect a peripheral device
(printer or PC) to or from the interface port.
Warmup Time
To deliver exact results, the balance must warm up
as listed below after initial connection to AC power
or after a relatively long power outage.
– Model LE26P…: at least 4 hours
– All other precision and analytical models:
at least 30 minutes
Only after this time will the balance have reached
the required operating temperature.
Using Verified Balances in Legal Metrology:
$
Allow the equipment to warm up for at least
24 hours after initial connection to AC power.
Leveling the Balance
Purpose:
– To compensate for unevenness at the place of
installation
Leveling Balances with a Weighing Capacity
up to 10 kg
Only the 2 front feet are adjusted to level the
balance.
●
Retract the two rear feet (only on models with
a rectangular weighing pan).
●
Turn the 2 front feet as shown in the diagram until
the air bubble is centered within the circle of the
level indicator.
> In most cases this will require several adjustment
steps.
●
For weighing heavy samples: Extend the 2 rear feet
until they touch the surface on which the balance
rests (only on models with a rectangular weighing
pan).
11