Internal Calibration/Adjustment
The menu code setting
1 9 3
must be selected in the Setup menu.
Inside the balance housing is a motori-
zed calibration weight which is applied
and removed automatically for internal
calibration.
●
Activate calibration: Press
q
> The built-in calibration weight
is applied automatically
> The balance is calibrated
> If “Calibrate, then auto adjust in one
operation" is selected in the Setup
menu, the balance is now adjusted
automatically.
> The internal calibration weight
is removed.
Calibration and Adjustment Sequence
In the Setup menu, you can configure
whether:
– calibration is always followed automati-
cally by adjustment (
1 10 1
; factory
setting), or
– you have the choice of ending the
sequence or starting adjustment after
calibration (
1 10 2
)
If no difference is determined between
nominal and actual weights, you can
end the calibration routine following
calibration.
Two keys are active at this point:
–
q
= start adjustment
–
c
= end the sequence