22
External Calibration
Step
Key (or instruction)
Display
1. Tare the balance.
w
0.0
g
2. Start calibration.
q
+
5000.0
g
CAL
G
3. Apply the prompted calibration
weight (in this example, 5000 g).
4. The balance is calibrated
–
0.2
g
(displayed only if menu
CAL
±
G
code
1 10 2
is set).
5. If the “Calibrate, then
Adjust
*
auto adjust" setting is
CAL
selected (
1 10 1
), the balance
is now adjusted automatically.
6. The calibration sequence is completed.
CC
CAL
7. After calibration/adjustment, the
+
5000.0
g
weight is displayed with weight unit
8. Remove the calibration weight.
0.0
g
* Adjust shown only if menu code
1 10 2
is set.
Important note:
Afterwards, do not perform internal calibration/adjustment again.