48
ISO/GLP-compliant Printout/Record
Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series. These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance manufacturer
– Balance model
– Balance serial number
– Software version number
– Identification number of the current
sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
The record is output to a Sartorius data
printer or a computer.
Settings
§
Setting menu codes for the printout
(see “Configuration"):
– ISO/GLP-compliant record after
calibration/adjustment only:
menu code
8 10 2
; or ISO/GLP-compliant
record always on: menu code
8 10 3
– Line format for printout: With data ID
codes - 22 characters: menu code
7 2 2
!
No ISO/GLP-compliant record is output
if any of the following settings are
configured:
menu code
6 1 4
,
6 1 5
,
6 1 6
(automatic printout) or
7 2 1
Function Keys
Press
r
to output header and first
measured value.
> Header is output the first time
r
is pressed
Press
v
to include output header
and reference data on automatic
printouts when an application
program is active
Close the application:
1) Output GLP footer:
Press
c
2) End application program:
Press
c