24
Overview
Specifications
Model
AZ214
AZ124
AZ64
Weighing capacity
g
210
120
60
Readability
mg
0.1
0.1
0.1
Tare range (subtractive)
g
210
120
60
Repeatability
<
±mg
0.2
0.2
0.2
Linearity
<
±mg
0.3
0.3
0.3
Operating temperature range
+10…+30 °C (50°F to 86°F)
Allowable ambient
operating temperature
+5… +40°C (41°F to 104°F)
Sensitivity drift
10…+30 °C (50°F–86°F)
>
±/K
2 · 10
–6
2 ·10
–6
2 ·10
–6
Response time (average)
s
3
3
3
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels
Display update (depends on the
filter level selected)
s
0.2–0.4
0.2–0.4
0.2–0.4
External calibration weight
g
200 (E2)
100 (E2)
50 (E2)
(of at least accuracy class)
lb
0.4
0.2
0.1
Net weight, approx.
kg
3.2
3.2
3.2
Pan size
mm
90
d
90
d
90
d
Weighing chamber height
mm
200
200
200
Dimensions (WxDxH)
mm
200
+
270
+
299
AC power source/
power requirements
AC adapter 230 V or 115 V, +15% to –20%
Frequency
48–60 Hz
AC power source,
direct current
V
10–20
Power consumption (average)
W
1
1
1
Hours of operation with the
YRB08Z rechargeable battery pack h
20
20
20
Summary of Contents for M-Power AZ124