9
Calibration/Adjustment
Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance,
– the balance is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed. Otherwise, the weight
required for calibration/adjustment is displayed (see “Accessories” for calibration weights).
You can use any of the following weight units to calibrate/adjust:
g, kg*, lb (menu code
1. 4.
x)
You can block calibration/adjustment of the balance:
– Select menu code
1. 5. 3
)
Example
Step
Key (or instruction) Display/Printout
1. Tare the balance
t
0.0
g
2. Begin calibration
t
>2 sec.
+
1000.0
Calibration weight is displayed
without weight unit (here: 1000 g)
3. Place the indicated calibration
1000.0
weight on the balance
After calibration, the calibration
+
1000.0
g
weight is displayed with wt. unit
4. Remove the calibration weight
0.0
g
* = not on models with a readability of 0.1 mg